4.5 Article

Improved Signal to Noise in Proton Spectroscopy of the Human Calf Muscle at 7 T Using Localized B(1) Calibration

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 63, 期 1, 页码 207-211

出版社

JOHN WILEY & SONS INC
DOI: 10.1002/mrm.22195

关键词

proton spectroscopy; B(1) mapping; 7 T; calf muscle; partial volume; coil

向作者/读者索取更多资源

Large variations of tip angle within a slice can lead to suboptimal pulse power optimization using standard techniques, which measure the average tip angle over a slice; this effect is especially pronounced at fields of 7 T and above. A technique was introduced that performed a volume-selective power optimization in less than 10 sec and automatically calibrates the radiofrequency pulses for subsequent spectroscopy scans. Using this technique, MR spectra were acquired in the human calf of seven healthy volunteers with a partial volume Tx/Rx coil. Increases in signal-to-noise ratio based upon the unsuppressed water signal between 22 +/- 5% and 166 +/- 42%, compared to spectra obtained with the conventional power calibration technique, were measured in different regions of the calf muscle. This new technique was able to measure the inhomogeneous radiofrequency field at 7 T and its use resulted in a considerable signal-to-noise ratio increase. Magn Reson Med 63:207-211, 2010. (C) 2009 Wiley-Liss, Inc.