期刊
MAGNETIC RESONANCE IN MEDICINE
卷 59, 期 3, 页码 655-660出版社
JOHN WILEY & SONS INC
DOI: 10.1002/mrm.21455
关键词
7 Tesla; GRAPPA; trabecular bone micro-architecture; cartilage
资金
- NIAMS NIH HHS [R01-AR 49701] Funding Source: Medline
The purpose of this work was to implement autocalibrating GRAPPA-based parallel imaging (PI) for in vivo high-resolution (HR) MRI of cartilage and trabecular bone micro-architecture at 7T and to evaluate its performance based on comparison of MR-derived morphology metrics between accelerated and conventional images and comparison of geometry factor measures between 3T and 7T. Using an eight channel coil array for trabecular MRI at the ankle, a higher maximum feasible acceleration (R) = 6 and lower geometry factor values than that at 3T were observed. The advantages of two-dimensional acceleration were also demonstrated. In knee cartilage and bone acquisitions, feasibility of PI with a dual-channel quadrature coil was investigated. Robust quantification of bone and cartilage metrics could be derived from accelerated ankle and knee acquisitions. PI can enhance the clinical feasibility of in vivo bone and cartilage HR-MRI for osteoporosis and osteoarthritis at 7T.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据