RF-related heating assessment of extracranial neurosurgical implants at 7 T

期刊

MAGNETIC RESONANCE IMAGING
卷 31, 期 6, 页码 1029-1034

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.mri.2012.10.025

关键词

7 T; Ultra-high-field MRI; MRI safety; Implants; SAR; Heating

资金

  1. Wright Center of Innovation in Biomedical Imaging
  2. [ODOD AGMT TECH 03-051]

向作者/读者索取更多资源

Purpose: The purpose was to evaluate radiofrequency (RF)-related heating of commonly used extracranial neurosurgical implants in 7-T magnetic resonance imaging (MRI). Materials and methods: Experiments were performed using a 7-T MR system equipped with a transmit/receive RF head coil. Four commonly used titanium neurosurgical implants were studied using a test procedure adapted from the American Society for Testing and Materials Standard F2182-11a. Implants (n = 4) were tested with an MRI turbo spin echo pulse sequence designed to achieve maximum RF exposure [specific absorption rate (SAR) level = 9.9 W/kg], which was further validated by performing calorimetry. Maximum temperature increases near each implant's surface were measured using fiberoptic temperature probes in a gelled-saline-filled phantom that mimicked the conductive properties of soft tissue. Measurement results were compared to literature data for patient safety. Results: The highest achievable phantom averaged SAR was determined by calorimetry to be 2.0 +/- 0.1 W/kg due to the highly conservative SAR estimation model used by this 7-T MR system. The maximum temperature increase at this SAR level was below 1.0 degrees C for all extracranial neurosurgical implants that underwent testing. Conclusion: The findings indicated that RF-related heating under the conditions used in this investigation is not a significant safety concern for patients with the particular extracranial neurosurgical implants evaluated in this study. (C) 2013 Elsevier Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据