4.3 Article

Immobilized bioactive agents onto polyurethane surface with heparin and phosphorylcholine group

期刊

MACROMOLECULAR RESEARCH
卷 21, 期 5, 页码 541-549

出版社

POLYMER SOC KOREA
DOI: 10.1007/s13233-013-1028-3

关键词

polyurethane; heparin; phosphorylcholine group; polyethylenimine; hemocompatibility

资金

  1. International Cooperation from Ministry of Science and Technology of China [2008DFA51170]
  2. Science and Technology Project of Tianjin Municipal Science and Technology Commission [08ZCKFSF03300]

向作者/读者索取更多资源

Heparin (HEP) and phosphorylcholine groups (PC) were grafted onto the polyurethane (PU) surface in order to improve biocompatibility and anticoagulant activity. After the surface grafting sites of PU were amplified with the primary amine groups of polyethylenimine (PEI), heparin was covalently linked onto the surface by the reaction between the amino group and the carboxyl group. PC groups were covalently immobilized on the PU-PEI surface through the reaction between the amino group and the aldehyde group of phosphorylcholine glyceraldehyde (PCGA). The surface density of primary amine groups was determined by a ninhydrin assay. The amino group density reached a maximum of 0.88 mu mol/cm(2) upon incorporation of 10 wt% PEI. The amount of heparin covalently immobilized on the PU-PEI surface was determined by the toluidine blue method. The grafting chemistry resulted in the comparatively dense immobilization of HEP (2.6 mu g/cm(2)) and PC to the PU-PEI surfaces. The HEP and PC modified surfaces were characterized by water uptake (PU 0.15 mg/cm(2), PU-PEI 3.54 mg/cm(2), PU-HEP 2.04 mg/cm(2), PU-PC 2.38 mg/cm(2)), water contact angle (PU 95.3A degrees, PU-PEI 34.0A degrees, PU-HEP 39.5A degrees, PU-PC 37.2A degrees), attenuated total reflection Fourier transform infrared spectroscopy (ATR-FTIR), X-ray photoelectron spectroscopy (XPS), atomic force microscopy (AFM), and scanning electron microscope (SEM). The results demonstrated that the PUPEI surface was successfully grafted with HEP and PC. The hydrophilicity and hemocompatibility of these grafted surfaces were significantly improved. These results suggested that the PU-HEP and PU-PC composite films are promising candidates for blood contacting tissue engineering.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据