4.7 Article

An Improved Index of Image Quality for Task-based Performance of CT Iterative Reconstruction across Three Commercial Implementations

期刊

RADIOLOGY
卷 275, 期 3, 页码 725-734

出版社

RADIOLOGICAL SOC NORTH AMERICA
DOI: 10.1148/radiol.15132091

关键词

-

资金

  1. National Institute of Standards and Technology
  2. Siemens Medical Solutions
  3. GE Healthcare

向作者/读者索取更多资源

Purpose: To develop and validate a metric of computed tomographic (CT) image quality that incorporates the noise texture and resolution properties of an image. Materials and Methods: Images of the American College of Radiology CT quality assurance phantom were acquired by using three commercial CT systems at seven dose levels with filtered back projection (FBP) and iterative reconstruction (IR). Image quality was characterized by the contrast-to-noise ratio (CNR) and a detectability index (d') that incorporated noise texture and spatial resolution. The measured CNR and d' were compared with a corresponding observer study by using the Spearman rank correlation coefficient to determine how well each metric reflects the ability of an observer to detect subtle lesions. Statistical significance of the correlation between each metric and observer performance was determined by using a Student t distribution; P values less than .05 indicated a significant correlation. Additionally, each metric was used to estimate the dose reduction potential of IR algorithms while maintaining image quality. Results: Across all dose levels, scanner models, and reconstruction algorithms, the d' correlated strongly with observer performance in the corresponding observer study (rho = 0.95; P < .001), whereas the CNR correlated weakly with observer performance (rho = 0.31; P = .21). Furthermore, the d' showed that the dose-reduction capabilities differed between clinical implementations (range, 12%-35%) and were less than those predicted from the CNR (range, 50%-54%). Conclusion: The strong correlation between the observer performance and the d' indicates that the d' is superior to the CNR for the evaluation of CT image quality. Moreover, the results of this study indicate that the d' improves less than the CNR with the use of IR, which indicates less potential for IR dose reduction than previously thought. (C)RSNA, 2015

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据