4.7 Article

Real-time feedback control of pH within microfluidics using integrated sensing and actuation

期刊

LAB ON A CHIP
卷 14, 期 6, 页码 1191-1197

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c3lc51205c

关键词

-

向作者/读者索取更多资源

We demonstrate a microfluidic system which applies engineering feedback principles to control the pH of a solution with a high degree of precision. The system utilizes an extended-gate ion-sensitive field-effect transistor (ISFET) along with an integrated pseudo-reference electrode to monitor pH values within a microfluidic reaction chamber. The monitored reaction chamber has an approximate volume of 90 nL. The pH value is controlled by adjusting the flow through two input channels using a pulse-width modulated signal applied to on-chip integrated valves. We demonstrate real-time control of pH through the feedback-controlled stepping of 0.14 pH increments in both the increasing and decreasing direction. The system converges to the pH setpoint within approximately 20 seconds of a step change. The integration of feedback theory into a microfluidic environment is a necessary step for achieving complete control over the microenvironment.