Lung assist device: development of microfluidic oxygenators for preterm infants with respiratory failure

期刊

LAB ON A CHIP
卷 13, 期 13, 页码 2641-2650

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c3lc41417e

关键词

-

资金

  1. Natural Sciences and Engineering Research Council of Canada (NSERC)
  2. Ontario Centres of Excellence (OCE)
  3. Ontario Ministry for Economic Development and Innovation through their Early Researcher Award program
  4. Canada Research Chairs Program

向作者/读者索取更多资源

This paper reports the development of microfluidic oxygenator (MFO) units designed for a lung assist device (LAD) for newborn infants. This device will be connected to the umbilical vessels like the natural placenta and provide gas exchange. The extracorporeal blood flow is only driven by the pressure difference between the umbilical artery and vein without the use of external pumps. The LAD is designed for use in ambient air (similar to 21% of 760 mmHg). The main focus of this paper is the presentation of the development of the MFO units testing various membrane materials with human blood to enhance gas exchange and in the design of fluidic inlets to lower the pressure drop across the oxygenator. Four different membranes, including thin film PDMS, porous PDMS, and two different pore size porous polycarbonate membranes are compared in this study. Among them, the microfluidic oxygenator with porous PDMS membrane has the highest gas exchange rate of 1.46 mu L min(-1) cm(2) for oxygen and 5.27 mu L min(-1) cm(2) for carbon dioxide and performs better than a commercial hollow fiber-based oxygenator by 367 and 233%, respectively. A new tapered inlet configuration was designed to reduce the pressure drop across the oxygenator and showed a further 57% improvement over the traditional perpendicular inlet configuration.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据