4.7 Article

Optical sensing system based on wireless paired emitter detector diode device and ionogels for lab-on-a-disc water quality analysis

期刊

LAB ON A CHIP
卷 12, 期 23, 页码 5069-5078

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c2lc40781g

关键词

-

资金

  1. EC FP7 People Programme, Science Foundation of Ireland [07/CE/I1147]
  2. Cytec Canada Inc
  3. Science Foundation Ireland [10/CE/B1821]

向作者/读者索取更多资源

This work describes the first use of a wireless paired emitter detector diode device (PEDD) as an optical sensor for water quality monitoring in a lab-on-a-disc device. The microfluidic platform, based on an ionogel sensing area combined with a low-cost optical sensor, is applied for quantitative pH and qualitative turbidity monitoring of water samples at point-of-need. The autonomous capabilities of the PEDD system, combined with the portability and wireless communication of the full device, provide the flexibility needed for on-site water testing. Water samples from local fresh and brackish sources were successfully analysed using the device, showing very good correlation with standard bench-top systems.