4.7 Article

Programmable large area digital microfluidic array with integrated droplet sensing for bioassays

期刊

LAB ON A CHIP
卷 12, 期 18, 页码 3305-3313

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c2lc40273d

关键词

-

资金

  1. Sharp Corporation

向作者/读者索取更多资源

We describe a new device concept for digital microfluidics, based on an active matrix electrowetting on dielectric (AM-EWOD) device. A conventional EWOD device is limited by the number of electrical connections that can be made practically, which restricts the number and type of droplet operations. In an AM-EWOD, the patterned electrodes of a conventional EWOD device are replaced by a thin film transistor (TFT) array, as found in a liquid crystal display (LCD), facilitating independent control of each electrode. The arrays can have many thousand individually addressable electrodes, are fully reconfigurable and can be programmed to support multiple simultaneous operations. Each element is 210 mu m 6 210 mu m in size and contains a circuit that measures the electrical impedance of the liquid above it. This is used to determine the presence and size of a droplet, a method that can improve assay reliability and accuracy. This sensor provides feedback, error detection and closed loop control of an assay sequence. We describe the design, fabrication and testing of a 64 x 64 format AM-EWOD device with impedance sensor functionality. A colorimetric assay is implemented on the device and used to measure glucose in human blood serum. Results are compared with the same assay performed on a microtitre plate.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据