4.7 Article

Real time quantitative amplification detection on a microarray: towards high multiplex quantitative PCR

期刊

LAB ON A CHIP
卷 12, 期 10, 页码 1897-1902

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c2lc20740k

关键词

-

向作者/读者索取更多资源

Quantitative real-time polymerase chain reaction (qrtPCR) is widely used as a research and diagnostic tool. Notwithstanding its many powerful features, the method is limited in the degree of multiplexing to about 6 due to spectral overlap of the available fluorophores. A new method is presented that allows quantitative amplification detection at higher multiplexing by the integration of amplification in solution and monitoring via hybridization to a microarray in real-time. This method does not require any manipulation of the PCR product and runs in a single closed chamber. Employing labeled primers, one of the main challenges is to measure surface signals against a high fluorescence background from solution. A compact, confocal scanner is employed, based on miniaturized optics from DVD technology and combined with a flat thermocycler for simultaneous scanning and heating. The feasibility of this method is demonstrated in singleplex with an analytical sensitivity comparable to routine qrtPCR.