4.7 Article

Design and dynamic characterization of single-stroke peristaltic PDMS micropumps

期刊

LAB ON A CHIP
卷 11, 期 2, 页码 336-342

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c0lc00023j

关键词

-

资金

  1. NIBIB NIH HHS [R01 EB009761] Funding Source: Medline
  2. NATIONAL INSTITUTE OF BIOMEDICAL IMAGING AND BIOENGINEERING [R01EB009761] Funding Source: NIH RePORTER

向作者/读者索取更多资源

In this paper, we present a monolithic PDMS micropump that generates peristaltic flow using a single control channel that actuates a group of different-sized microvalves. An elastomeric microvalve design with a raised seat, which improves bonding reliability, is incorporated into the micropump. Pump performance is evaluated based on several design parameters-size, number, and connection of successive microvalves along with control channel pressure at various operating frequencies. Flow rates ranging 0-5.87 mu L min(-1) were observed. The micropump design demonstrated here represents a substantial reduction in the number of/real estate taken up by the control lines that are required to run a peristaltic pump, hence it should become a widespread tool for parallel fluid processing in high-throughput microfluidics.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据