期刊
LAB ON A CHIP
卷 10, 期 3, 页码 281-290出版社
ROYAL SOC CHEMISTRY
DOI: 10.1039/b914183a
关键词
-
类别
资金
- European Commission [026554]
- Steinmaur Foundation, Liechtenstein
We present the design, fabrication, and characterisation of an array of optical slot-waveguide ring resonator sensors, integrated with microfluidic sample handling in a compact cartridge, for multiplexed real-time label-free biosensing. Multiplexing not only enables high throughput, but also provides reference channels for drift compensation and control experiments. Our use of alignment tolerant surface gratings to couple light into the optical chip enables quick replacement of cartridges in the read-out instrument. Furthermore, our novel use of a dual surface-energy adhesive film to bond a hard plastic shell directly to the PDMS microfluidic network allows for fast and leak-tight assembly of compact cartridges with tightly spaced fluidic interconnects. The high sensitivity of the slot-waveguide resonators, combined with on-chip referencing and physical modelling, yields a volume refractive index detection limit of 5 x 10(-6) refractive index units (RIUs) and a surface mass density detection limit of 0.9 pg mm(-2), to our knowledge the best reported values for integrated planar ring resonators.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐