期刊
LAB ON A CHIP
卷 9, 期 11, 页码 1625-1630出版社
ROYAL SOC CHEMISTRY
DOI: 10.1039/b819688e
关键词
-
类别
资金
- CCMX-MMNS
- European Project 'Good Food' [FP6-IST-1-508774-IP]
- OFFT
A multiplexed immunoassay-based antibiotic sensing device integrated in a lab-on-a-chip format is described. The approach is multidisciplinary and involves the convergent development of a multi-antibiotic competitive immunoassay based on sensitive wavelength interrogated optical sensor (WIOS) technology and a polymer-based self-contained microfluidic cartridge. Immunoassay solutions are pressure-driven through external and concerted actuation of a single syringe pump and multiposition valve. Moreover, the use of a novel photosensitive material in a 'one step' fabrication process allowed the rapid fabrication of microfluidic components and interconnection port simultaneously. Pre-filled microfluidic cartridges were used as binary response rapid tests for the simultaneous detection of three antibiotic families - sulfonamides, fluoroquinolones and tetracyclines - in raw milk. For test interpretation, any signal lower than the threshold value obtained for the corresponding Maximum Residue Limit (MRL) concentration (100 mu g L(-1)) was considered negative for a given antibiotic. The reliability of the multiplexed detection system was assessed by way of a validation test carried out on a series of six blind milk samples. A test accuracy of 95% was calculated from this experiment. The whole immunoassay procedure is fast (less than 10 minutes) and easy to handle (automated actuation).
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐