期刊
LAB ON A CHIP
卷 8, 期 12, 页码 2113-2120出版社
ROYAL SOC CHEMISTRY
DOI: 10.1039/b811169c
关键词
-
类别
资金
- DARPA [AF F30602-00-1-0569]
This paper presents the development of an easy-to-handle and disposable clinical diagnostic lab-on-achip using fully integrated plastic microfluidic components, which has the sampling/identifying capability to make fast and reliable measurements of metabolic parameters from human whole blood. A smart and functional lab-on-a-chip cartridge, which incorporates a full on-chip auto-calibration function for in the field applications, has been developed, and then fully characterized using a portable analyzer (3(1/4)'' x 5 '' x 1 '') with multi-analyte detection capability. In addition, several new approaches in realizing smart and functional lab-on-a-chips on polymer have been adopted, which include the pinch valve for automatic fluidic sealing, a by-pass channel as the sampling indicator, and a robust connector design for long analyzer lifetimes. Metabolic parameters such as glucose, lactate, and partial oxygen from human whole blood have been successfully measured using the functional polymer lab-on-a-chips and the portable analyzer developed in this work.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据