4.5 Article

A mechanized and standardized pivot shifter: technical description and first evaluation

期刊

出版社

SPRINGER
DOI: 10.1007/s00167-010-1289-8

关键词

Pivot shift; CAOS; ACL

向作者/读者索取更多资源

The pivot shift test (PST) is a complex, multiplanar maneuver used to assess rotatory instability of the knee. The grading is subjective due to the broad range of examination techniques and lack of tibiofemoral motion quantification. The goal of this study was to develop and evaluate a mechanized device for quantitative assessment of the PST. We constructed a mechanized pivot shifter (MPS). In five cadaveric hip-to-toes specimens, the anterior cruciate ligament was resected. We used a surgical navigation system for acquisition of the tibiofemoral motion path during the PST. Two sets of measurements were obtained for the MPS and for two examiners performing the manual technique. Mean lateral compartment translation magnitudes for each MPS measurement were 13.5 mm (sigma = 6.7) and 13.6 mm (sigma = 6.7). For examiner 1, 14.9 mm (sigma = 6.5) and 15.7 mm (sigma = 6.3). For examiner 2, 16.9 mm (sigma = 6.3) and 16.1 mm (sigma = 5.2). Differences were not significant (n.s.). The MPS had narrower limits of agreement than both examiner 1 and examiner 2. The MPS demonstrated no significant differences in the tibiofemoral translation magnitudes compared to the manual technique. It resulted in better test-retest reliability and more consistent measurements of tibiofemoral translation when compared to manual PST. The high repeatability factor conferred by the MPS is a clinical advantage.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据