Several studies have stressed the importance of dialysis time in the removal of uremic retention solutes. To further investigate this, nine stable chronic hemodialysis patients were dialyzed for 4, 6, or 8 h processing the same total blood and dialysate volume by the Genius system and high-flux FX80 dialyzers. Inlet blood and outlet dialysate were analyzed for urea, creatinine, phosphorus, and beta 2-microglobulin at various times. Total solute removal, dialyzer extraction ratios, and total cleared volumes were significantly larger during prolonged dialysis for urea, creatinine, phosphorus, and beta 2-microglobulin. Reduction ratios increased progressively, except for phosphate and beta 2-microglobulin, where the ratios remained constant after 2 h. In contrast, no significant difference was found for the reduction ratios of all solutes and Kt/V-urea between the three different sessions. With longer dialyses, solutes are efficiently removed from the deeper compartments of the patient's body. Our study shows that care must be taken when using Kt/V-urea or reduction ratios as the only parameters to quantify dialysis adequacy.