4.6 Article

Refinement of the cutoff values of the HemosIL AcuStar assay for the detection of anticardiolipin and anti-beta2 glycoprotein-1 antibodies

期刊

JOURNAL OF THROMBOSIS AND HAEMOSTASIS
卷 12, 期 12, 页码 2034-2037

出版社

WILEY
DOI: 10.1111/jth.12732

关键词

anticardiolipin antibody; antiphospholipid antibodies; antiphospholipid antibody syndrome; beta 2-Glycoprotein I; chemiluminescence

向作者/读者索取更多资源

BackgroundThe HemosIL AcuStar antiphospholipid assay (Instrumentation Laboratory, Bedford, MA, USA) is a fully automated assay using chemiluminescent technology for the detection of anticardiolipin and anti-beta(2) glycoprotein-1 antibodies. This assay showed excellent agreement between results of different laboratories. The cutoff values to define positivity were calculated in 250 healthy blood bank donors but were associated with large confidence intervals (CIs). ObjectiveThe objective of this study was to more precisely determine the cutoff values of the HemosIL AcuStar antiphospholipid assay by increasing the number of healthy blood bank donors through a multicenter study and by applying a normalization procedure of the distribution of each antibody. MethodsFive laboratories participated to this study, allowing the inclusion of 626 samples. We used a Box-Cox power transformation method to normalize the distribution and calculate the 99th percentile and the corresponding 95%CI for each antibody. ResultsThe revised cutoff values were overall lower than those initially calculated with more stringent CIs and yielded a 4.2% increase in sensitivity with a 2.7% decrease in specificity regarding thrombotic events or obstetric complications. ConclusionsWe provide refined cutoff values for the detection of anticardiolipin and anti-beta(2) glycoprotein-1 antibodies with the HemosIL AcuStar Antiphospholipid assay that should be preferred for routine use.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据