4.6 Article

Computer simulating a clinical trial of a load-bearing implant: An example of an intramedullary prosthesis

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jmbbm.2011.06.005

关键词

Simulated clinical trials; Intramedullary fixation; Stochastic model; Finite element analysis; Mechanobiology

资金

  1. Science Foundation Ireland

向作者/读者索取更多资源

Computational modelling is becoming ever more important for obtaining regulatory approval for new medical devices. An accepted approach is to infer performance in a population from an analysis conducted for an idealised or 'average' patient; we present here a method for predicting the performance of an orthopaedic implant when released into a population-effectively simulating a clinical trial. Specifically we hypothesise that an analysis based on a method for predicting the performance in a population will lead to different conclusions than an analysis based on an idealised or 'average' patient. To test this hypothesis we use a finite element model of an intramedullary implant in a bone whose size and remodelling activity is different for each individual in the population. We compare the performance of a low Young's modulus implant (E = 20 GPa) to one with a higher Young's modulus (200 GPa). Cyclic loading is applied and failure is assumed when the migration of the implant relative to the bone exceeds a threshold magnitude. The analysis for an idealised of 'average' patient predicts that the lower modulus device survives longer whereas the analysis simulating a clinical trial predicts no statistically-significant tendency (p = 0.77) for the low modulus device to perform better. It is concluded that population-based simulations of implant performance - simulating a clinical trial - present a very valuable opportunity for more realistic computational pre-clinical testing of medical devices. (C) 2011 Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据