4.3 Article

Simultaneous Determination of Aluminum, Cadmium and Lead in Whole Blood by Simultaneous Atomic Absorption Spectrometry with Oxygen Charring

期刊

JOURNAL OF THE CHINESE CHEMICAL SOCIETY
卷 56, 期 1, 页码 135-141

出版社

WILEY-V C H VERLAG GMBH
DOI: 10.1002/jccs.200900019

关键词

Aluminum; Cadmium; Lead; Whole blood; Simultaneous atomic absorption spectrometry; Oxygen charring

资金

  1. National Science Council of Taiwan [NSC 92-2119-M-242-001]

向作者/读者索取更多资源

A method for the simultaneous determination of aluminum (A I), cadmium (Cd) and lead (Pb) in whole blood has been developed by using simultaneous atomic absorption spectrometry (SIMAAS) with oxygen charring. The optimized conditions for the simultaneous determination of AI, Cd and Pb were obtained in the presence of palladium (Pd) as the chemical modifier, using 600 degrees C and 2400 degrees C as the pyrolysis and the atomization temperature, respectively. The whole blood samples were diluted 1+5 (v/v) directly with 0.1% (v/v) Triton X-100. Oxygen was employed to eliminate the interference of carbonaceous residues in the charring step before pyrolysis. The calibration curves were carried out with aqueous standard solutions and the linear ranges were 0-40 ng mL(-1), 0-4 ng mL(-1) and 0-40 ng mL(-1) for Al, Cd and Pb, respectively. The detection limits were 0.96 ng mL(-1) (19.2 pg) for Al, 0.03 ng mL(-1) (0.6 pg) for Cd and 0.60 ng mL(-1) (12.0 pg) for Pb. The spiked recoveries of Al, Cd and Pb in whole blood were 98.0%, 100.0% and 101.7%, respectively. The accuracy of the proposed method was evaluated with the analysis of a whole blood certified reference material (Seronorm, level 2). The found concentrations were in agreement with the recommended values. The proposed method has been successfully applied to the simultaneous determination of Al, Cd and Pb in whole blood of healthy volunteers before and after eating barbecued foods.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据