4.6 Article

Simple Experimental Methods for Determining the Apparent Focal Shift in a Microscope System

期刊

PLOS ONE
卷 10, 期 8, 页码 -

出版社

PUBLIC LIBRARY SCIENCE
DOI: 10.1371/journal.pone.0134616

关键词

-

资金

  1. National Science Foundation [PHY-0844466]
  2. National Institutes of Health [P50 GM071508]

向作者/读者索取更多资源

Three-dimensional optical microscopy is often complicated by a refractive index mismatch between the sample and objective lens. This mismatch causes focal shift, a difference between sample motion and focal-plane motion, that hinders the accuracy of 3D reconstructions. We present two methods for measuring focal shift using fluorescent beads of different sizes and ring-stained fluorescent beads. These simple methods are applicable to most situations, including total internal reflection objectives and samples very close to the interface. For distances 0-1.5 mu m into an aqueous environment, our 1.49-NA objective has a relative focal shift of 0.57 +/- 0.02, significantly smaller than the simple n(2)/n(1) approximation of 0.88. We also expand on a previous sub-critical angle theory by means of a simple polynomial extrapolation. We test the validity of this extrapolation by measuring the apparent focal shift in samples where the refractive index is between 1.33 and 1.45 and with objectives with numerical apertures between 1.25 and 1.49.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据