Magnetic Resonance Imaging at 1.5-T in Patients With Implantable Cardioverter-Defibrillators

期刊

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.jacc.2009.04.050

关键词

magnetic resonance imaging; implantable cardioverter-defibrillator; safety

向作者/读者索取更多资源

Objectives Our aim was to establish and evaluate a strategy for safe performance of magnetic resonance imaging (MRI) at 1.5-T in patients with implantable cardioverter-defibrillators (ICDs). Background Expanding indications for ICD placement and MRI becoming the imaging modality of choice for many indications has created a growing demand for MRI in ICD patients, which is still considered an absolute contraindication. Methods Non-pacemaker-dependent ICD patients with a clinical need for MRI were included in the study. To minimize radiofrequency-related lead heating, the specific absorption rate was limited to 2 W/kg. ICDs were reprogrammed pre-MRI to avoid competitive pacing and potential pro-arrhythmia: 1) the lower rate limit was programmed as low as reasonably achievable; and 2) arrhythmia detection was programmed on, but therapy delivery was programmed off. Patients were monitored using electrocardiography and pulse oximetry. All ICDs were interrogated before and after the MRI examination and after 3 months, including measurement of pacing capture threshold, lead impedance, battery voltage, and serum troponin I. Results Eighteen ICD patients underwent a total of 18 MRI examinations at 1.5-T; all examinations were completed safely. All ICDs could be interrogated and reprogrammed normally post-MRI. No significant changes of pacing capture threshold, lead impedance, and serum troponin I were observed. Battery voltage decreased significantly from pre-to post-MRI. In 2 MRI examinations, oversensing of radiofrequency noise as ventricular fibrillation occurred. However, no attempt at therapy delivery was made. Conclusions MRI of non-pacemaker-dependent ICD patients can be performed with an acceptable risk/benefit ratio under controlled conditions by taking both MRI-and pacemaker-related precautions. (Implantable Cardioverter Defibrillators and Magnetic Resonance Imaging of the Heart at 1.5-Tesla; NCT00356239) (J Am Coll Cardiol 2009; 54: 549-55) (C) 2009 by the American College of Cardiology Foundation

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据