Inaccuracy of Venous Point-of-Care Glucose Measurements in Critically Ill Patients: A Cross-Sectional Study

期刊

PLOS ONE
卷 10, 期 6, 页码 -

出版社

PUBLIC LIBRARY SCIENCE
DOI: 10.1371/journal.pone.0129568

关键词

-

资金

  1. Instituto de Ensino e Pesquisa (IIEP) - Hospital Israelita Albert Einstein [80-05]

向作者/读者索取更多资源

Introduction Current guidelines and consensus recommend arterial and venous samples as equally acceptable for blood glucose assessment in point-of-care devices, but there is limited evidence to support this recommendation. We evaluated the accuracy of two devices for bedside point-of-care blood glucose measurements using arterial, fingerstick and catheter venous blood samples in ICU patients, and assessed which factors could impair their accuracy. Methods 145 patients from a 41-bed adult mixed-ICU, in a tertiary care hospital were prospectively enrolled. Fingerstick, central venous (catheter) and arterial blood (indwelling catheter) samples were simultaneously collected, once per patient. Arterial measurements obtained with Precision PCx, and arterial, fingerstick and venous measurements obtained with Accu-chek Advantage II were compared to arterial central lab measurements. Agreement between point-of-care and laboratory measurements were evaluated with Bland-Altman, and multiple linear regression models were used to investigate interference of associated factors. Results Mean difference between Accu-chek arterial samples versus central lab was 10.7 mg/dL (95% LA -21.3 to 42.7 mg/dL), and between Precision PCx versus central lab was 18.6 mg/dL (95% LA -12.6 to 49.5 mg/dL). Accu-chek fingerstick versus central lab arterial samples presented a similar bias (10.0 mg/dL) but a wider 95% LA (-31.8 to 51.8 mg/dL). Agreement between venous samples with arterial central lab was the poorest (mean bias 15.1 mg/dL; 95% LA -51.7 to 81.9). Hyperglycemia, low hematocrit, and acidosis were associated with larger differences between arterial and venous blood measurements with the two glucometers and central lab. Vasopressor administration was associated with increased error for fingerstick measurements. Conclusions Sampling from central venous catheters should not be used for glycemic control in ICU patients. In addition, reliability of the two evaluated glucometers was insufficient. Error with Accu-chek Advantage II increases mostly with central venous samples. Hyperglycemia, lower hematocrit, acidosis, and vasopressor administration increase measurement error.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据