4.0 Article

Factors Influencing Individual Variation in Perceptual Directional Microphone Benefit

期刊

JOURNAL OF THE AMERICAN ACADEMY OF AUDIOLOGY
卷 24, 期 10, 页码 955-968

出版社

AMER ACAD AUDIOLOGY
DOI: 10.3766/jaaa.24.10.7

关键词

Directional benefit; directional microphone; directivity index; hearing aid; in situ measurements; speech reception threshold in noise; test-retest measurements

资金

  1. HEARing Cooperative Research Centre

向作者/读者索取更多资源

Background: Large variations in perceptual directional microphone benefit, which far exceed the variation expected from physical performance measures of directional microphones, have been reported in the literature. The cause for the individual variation has not been systematically investigated. Purpose: To determine the factors that are responsible for the individual variation in reported perceptual directional benefit. Research Design: A correlational study. Physical performance measures of the directional microphones obtained after they had been fitted to individuals, cognitive abilities of individuals, and measurement errors were related to perceptual directional benefit scores. Study Sample: Fifty-nine hearing-impaired adults with varied degrees of hearing loss participated in the study. Data Collection and Analysis: All participants were bilaterally fitted with a Motion behind-the-ear device (500 M, 501 SX, or 501 P) from Siemens according to the National Acoustic Laboratories' non-linear prescription, version two (NAL-NL2). Using the Bamford-Kowal-Bench (BKB) sentences, the perceptual directional benefit was obtained as the difference in speech reception threshold measured in babble noise (SRTn) with the devices in directional (fixed hypercardioid) and in omnidirectional mode. The SRTn measurements were repeated three times with each microphone mode. Physical performance measures of the directional microphone included the angle of the microphone ports to loudspeaker axis, the frequency range dominated by amplified sound, the in situ signal-to-noise ratio (SNR), and the in situ threedimensional, articulation-index weighted directivity index (3D AI-DI). The cognitive tests included auditory selective attention, speed of processing, and working memory. Intraparticipant variation on the repeated SRTn's and the interparticipant variation on the average SRTn were used to determine the effect of measurement error. A multiple regression analysis was used to determine the effect of other factors. Results: Measurement errors explained 52% of the variation in perceptual directional microphone benefit (95% confidence interval [Cl]: 34-78%), while another 37% of variation was explained primarily by the physical performance of the direcfional microphones after they were fitted to individuals. The most contributing factor was the in situ 3D AI-DI measured across the low frequencies. Conclusions: Repeated SRTn measurements are needed to obtain a reliable indication of the perceptual directional benefit in an individual. Further, to obtain optimum benefit from directional microphones, the effectiveness of the microphones should be maximized across the low frequencies.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据