4.2 Article Proceedings Paper

Capturing dynamics with Eiger, a fast-framing X-ray detector

期刊

JOURNAL OF SYNCHROTRON RADIATION
卷 19, 期 -, 页码 1001-1005

出版社

INT UNION CRYSTALLOGRAPHY
DOI: 10.1107/S0909049512035972

关键词

detector; single photon counting; X-ray diffraction; X-ray photon correlation spectroscopy

向作者/读者索取更多资源

Eiger is the next-generation single-photon-counting pixel detector following the widely used Pilatus detector. Its smaller pixel size of 75 mu m x 75 mu m, higher frame rate of up to 22 kHz, and practically zero dead-time (similar to 4 mu s) between exposures will further various measurement methods at synchrotron sources. In this article Eiger's suitability for X-ray photon correlation spectroscopy (XPCS) is demonstrated. By exploiting its high frame rate, complementary small-angle X-ray scattering (SAXS) and XPCS data are collected in parallel to determine both the structure factor and collective diffusion coefficient of a nano-colloid suspension. For the first time, correlation times on the submillisecond time scale are accessible with a large-area pixel detector.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据