4.5 Article

Evaluation of a hydrogel based breast biopsy marker (HydroMARK®) as an alternative to wire and radioactive seed localization for non-palpable breast lesions

期刊

JOURNAL OF SURGICAL ONCOLOGY
卷 105, 期 6, 页码 591-594

出版社

WILEY
DOI: 10.1002/jso.22146

关键词

breast surgery; preoperative localization; ultrasound

向作者/读者索取更多资源

Background and Objectives HydroMARK (R) is a newly available biopsy marker for image-guided needle biopsies of non-palpable breast lesions. Objective was to determine if the marker could be utilized independently for lesion localization using intra-operative ultrasound alone. Methods: A single institution retrospective review identified patients who underwent surgical excision of breast lesions after placement of the HydroMARK1. Endpoints included intra-operative visualization of the marker, successful excision of the lesion, and presence of the marker on specimen radiograph. Results: The study included 31 lesions in 25 patients. Twenty-nine (93.6%) HydroMARKSs1 were adequately visualized by intra-operative ultrasound. Intra-operative ultrasound without pre-operative placement of a localizing device was successful for localization in six cases (19.4%). Intra-operative difficulties were encountered in 16 of 31 (51.6%) procedures. This included either extrusion of the marker when the biopsy tract was transected in 14 (45.2%) cases or migration of the marker prior to the procedure in two (6.4%) cases. The marker was visualized on specimen radiograph in 15 (48.4%) cases. Conclusions: While intraoperative sonographic visibility was excellent, a large number of excisions were associated with extrusion of the marker. Modifications are needed to improve acceptability of this marker for intra-operative localization independent of pre-operative wire or seed localization. J. Surg. Oncol. 2012; 105: 591-594. (C) 2011 Wiley Periodicals, Inc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据