4.1 Article

Modified Screen-Printed Electrode for Potentiometric Determination of Copper(II) in Water Samples

期刊

JOURNAL OF SOLUTION CHEMISTRY
卷 42, 期 6, 页码 1336-1354

出版社

SPRINGER/PLENUM PUBLISHERS
DOI: 10.1007/s10953-013-0030-x

关键词

Screen printed ion selective electrode; Modified carbon paste electrode; Copper sulfate; Potentiometric titration; Water samples

向作者/读者索取更多资源

Modified screen printed (SPE) and carbon paste electrodes (CPE) with phenanthroline-tetraphenyl borate ionophore [Phen:TPB] were fabricated for the determination of copper(II). The modified electrodes have linear responses over a wide concentration range (1 x 10(-6)-1 x 10(-2) mol center dot L-1) of copper(II) ion at 25 A degrees C with divalent cationic slopes of 29.85 +/- A 0.58 and 29.45 +/- A 0.81 mV center dot decade(-1) and exhibit a detection limit of 1 x 10(-6) mol center dot L-1 for SPE and CPE. The selectivity coefficient was measured using the match potential method in acetate buffer of pH = 4.2. The modified SPE and CPE sensors show high selectivity and sensitivity for determination of copper(II) and also show stable and reproducible response over a period of five and three months for SPE and CPE sensors, respectively. This method can be used for determination of copper(II) in water, soil, plant and fish tissue samples and the results obtained agreed with those obtained with atomic absorption spectrometer (AAS).

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据