4.5 Review

Resolving power in liquid chromatography: A trade-off between efficiency and analysis time

期刊

JOURNAL OF SEPARATION SCIENCE
卷 42, 期 1, 页码 38-50

出版社

WILEY-V C H VERLAG GMBH
DOI: 10.1002/jssc.201800891

关键词

kinetic performance; kinetic plot; ultra high performance liquid chromatography; ultra high pressure liquid chromatography; van Deemter curve

资金

  1. Research Foundation Flanders (FWO) [G025916N, 12J6517N, 1508914N]

向作者/读者索取更多资源

yThis review describes chromatographic dispersion and different plate-height models frequently used to assess the chromatographic performance of ultra-high-pressure liquid chromatography column technology. Furthermore, different performance indices, including the resolution, the separation impedance, and kinetic plots are discussed allowing to quantify and visualize the resolving power in liquid chromatography. The construction of kinetic plots is explained, and different visualization approaches are highlighted. Finally, key instrument and column-technology developments to advance the kinetic performance limits are discussed and selected state-of-the-art applications are highlighted.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据