4.5 Article

Determination of sulfonamides in milk samples by HPLC with amperometric detection using a glassy carbon electrode modified with multiwalled carbon nanotubes

期刊

JOURNAL OF SEPARATION SCIENCE
卷 37, 期 4, 页码 382-389

出版社

WILEY-V C H VERLAG GMBH
DOI: 10.1002/jssc.201301011

关键词

Amperometric detection; Carbon nanotubes; Liquid chromatography; Modified electrodes; Sulfonamides

资金

  1. Spanish Ministry of Science and Innovation [CTQ2010-15027]

向作者/读者索取更多资源

A sensitive and accurate method for determining five sulfonamides based on HPLC with amperometric detection and using a glassy carbon electrode modified with multiwalled carbon nanotubes is proposed. Optimal conditions for the quantitative separation of selected sulfonamides were studied, and glassy carbon electrodes with and without modification with carbon nanotubes were systematically investigated as electrodic materials. Statistical analysis of the obtained results demonstrated that these modified electrodes achieved considerably better stability and sensitivity than the conventional unmodified ones. Detection limits were in the 1.2-6.0 ng/mL range. The usefulness of the method was demonstrated by the analysis of milk samples, taking into account the European legislation on residues in food products, following both a screening method to classify the samples and a confirmation method to provide more detailed information in the case of positive samples.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据