The application of novel 1.7 μm ethylene bridged hybrid particles for hydrophilic interaction chromatography

期刊

JOURNAL OF SEPARATION SCIENCE
卷 31, 期 9, 页码 1511-1518

出版社

WILEY-V C H VERLAG GMBH
DOI: 10.1002/jssc.200700673

关键词

ethylene bridged hybrid particle; ESI-MS sensitivity; hybrid particles; hydrophilic interaction chromatography; polar retention

向作者/读者索取更多资源

An un-derivatized 1.7 mu m ethylene bridged hybrid (BEH) particle was evaluated for its utility in retaining polar species in hydrophilic interaction chromatography (HILIC), and was compared to a 3 mu m un-derivatized silica material. Retentivity as a function of mobile phase pH, polar modifier and ACN content was examined. Also, the efficiency of the two particle substrates was compared by plotting HETP vs. linear velocity. Improved chemical resistance of the un-derivatized BEH particle was compared to un-derivatized silica at pH 5, demonstrating no performance deterioration over the course of 2000 injections for the BEH particle, while the silica particle deteriorated rapidly after 800 injections. Lastly, ESI-MS sensitivity as a function of particle size and separation mode was demonstrated. A 2.2 to 4.7-times higher ESI-MS response was observed on the 1.7 mu m particle compared to the 3 mu m particle, whereas a 5.6 to 8.8-times higher ESI-MS response was observed using HILIC as when compared to traditional RP chromatography.