4.5 Article

Isotopic analysis of uranium in NIST SRM glass by femtosecond laser ablation MC-ICPMS

期刊

出版社

SPRINGER
DOI: 10.1007/s10967-012-2218-8

关键词

Laser ablation; Multicollector mass spectroscopy; ICPMS; NIST glass; Uranium; Isotope ratio

资金

  1. United States Department of Energy, National Nuclear Security Administration, Office of Nonproliferation Verification Research and Development [NA-22]

向作者/读者索取更多资源

We employed femtosecond Laser Ablation Multicollector Inductively Coupled Mass Spectrometry for the determination of uranium isotope ratios in a series of standard reference material glasses (NIST 610, 612, 614, and 616). The uranium in this series of SRM glasses is a combination of isotopically natural uranium in the materials used to make the glass matrix and isotopically depleted uranium added to increase the uranium elemental concentration across the series. Results for NIST 610 are in excellent agreement with literature values. However, other than atom percent U-235, little information is available for the remaining glasses. We present atom percent and isotope ratios for U-234, U-235, U-236, and U-238 for all four glasses. Our results show deviations from the certificate values for the atom percent U-235, indicating the need for further examination of the uranium isotopes in NIST 610-616.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据