期刊
JOURNAL OF PROTEOME RESEARCH
卷 7, 期 4, 页码 1640-1649出版社
AMER CHEMICAL SOC
DOI: 10.1021/pr700675j
关键词
phosphoproteome analysis; MS2/MS3; target-decoy search; automatic validation
Manual checking is commonly employed to validate the phosphopeptide identifications from database searching of tandem mass spectra. It is very time-consuming and labor intensive as the number of phosphopeptide identifications increases greatly. In this study, a simple automatic validation approach was developed for phosphopeptide identification by combining consecutive stage mass spectrometry data and the target-decoy database searching strategy. Only phosphopeptides identified from both MS2 and its corresponding MS3 were accepted for further filtering, which greatly improved the reliability in phosphopepticle identification. Before database searching, the spectra were validated for charge state and neutral loss peak intensity, and then the invalid MS2/MS3 spectra were removed, which greatly reduced the database searching time. It was found that the sensitivity was significantly improved in MS2/MS3 strategy as the number of identified phosphopeptides was 2.5 times that obtained by the conventional filter-based MS2 approach. Because of the use of the target-decoy database, the false-discovery rate (FDR) of the identified phosphopeptides could be easily determined, and it was demonstrated that the determined FDR can precisely reflect the actual FDR without any manual validation stage.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据