Evaluation of a novel portable x-ray fluorescence screening tool for detection of arsenic exposure

期刊

PHYSIOLOGICAL MEASUREMENT
卷 36, 期 12, 页码 2443-2459

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0967-3334/36/12/2443

关键词

arsenic; x-ray fluorescence; diagnostic test; inductively coupled plasma-mass spectrometry

资金

  1. Health Canada
  2. Canadian Institutes of Health Research [CDA-66534]
  3. Natural Sciences and Engineering Research Council of Canada
  4. Canada Research Chair program

向作者/读者索取更多资源

A new portable x-ray fluorescence (XRF) screening tool was evaluated for its effectiveness in arsenic (As) quantification in human finger and toe nails (n = 58). Nail samples were measured for total As concentration by XRF and inductively coupled plasma-mass spectrometry (ICP-MS). Using concordance correlation coefficient (CCC), kappa, diagnostic sensitivity (Sn) and specificity (Sp), and linear regression analyses, the concentration of As measured by XRF was compared to ICP-MS. The CCC peaked for scaled values of fingernail samples, at 0.424 (95% CI: 0.065-0.784). The largest kappa value, 0.400 (95% CI: -0.282-1.000), was found at a 1.3 mu g g(-1) cut-off concentration, for fingernails only, and the largest kappa at a clinically relevant cut-off concentration of 1.0 mu g g(-1) was 0.237 (95% CI: -0.068-0.543), again in fingernails. Analyses generally showed excellent XRF Sn (up to 100%, 95% CI: 48-100%), but low Sp (up to 30% for the same analysis, 95% CI: 14-50%). Portable XRF shows some potential for use as a screening tool with fingernail samples. The difference between XRF and ICP-MS measurements decreased as sample mass increased to 30 mg. While this novel method of As detection in nails has shown relatively high agreement in some scenarios, this portable XRF is not currently considered suitable as a substitute for ICP-MS.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据