4.6 Article

Design factors of intravascular dual frequency transducers for super-harmonic contrast imaging and acoustic angiography

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 60, 期 9, 页码 3441-3457

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/60/9/3441

关键词

ultrasound transducer; intravascular ultrasound imaging; acoustic angiography; contrast imaging; super-harmonic imaging; dual frequency transducer; microbubble

资金

  1. NIH [R01EB015508, R01CA170665, P41EB02182]

向作者/读者索取更多资源

Imaging of coronary vasa vasorum may lead to assessment of the vulnerable plaque development in diagnosis of atherosclerosis diseases. Dual frequency transducers capable of detection of microbubble superharmonics have shown promise as a new contrast-enhanced intravascular ultrasound (CE-IVUS) platform with the capability of vasa vasorum imaging. Contrast-to-tissue ratio (CTR) in CE-IVUS imaging can be closely associated with low frequency transmitter performance. In this paper, transducer designs encompassing different transducer layouts, transmitting frequencies, and transducer materials are compared for optimization of imaging performance. In the layout selection, the stacked configuration showed superior super-harmonic imaging compared with the interleaved configuration. In the transmitter frequency selection, a decrease in frequency from 6.5 MHz to 5 MHz resulted in an increase of CTR from 15 dB to 22 dB when receiving frequency was kept constant at 30 MHz. In the material selection, the dual frequency transducer with the lead magnesium niobate-lead titanate (PMN-PT) 1-3 composite transmitter yielded higher axial resolution compared to single crystal transmitters (70 mu m compared to 150 mu m pulse length). These comparisons provide guidelines for the design of intravascular acoustic angiography transducers.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据