Reliability of computer-assisted method transfer between several column dimensions packed with 1.3-5 μm core-shell particles and between various instruments

期刊

出版社

ELSEVIER
DOI: 10.1016/j.jpba.2014.01.037

关键词

Method transfer; Method development; DryLab; Modelling software; Core-shell particles

向作者/读者索取更多资源

In this contribution, the possibility to automatically transfer RPLC methods between different column dimensions and instruments was evaluated using commercial modelling software. The method transfer reliability was tested with loratadine and its 7 related pharmacopeia] impurities. In this study, state-of-the-art columns packed with superficially porous particles of 5, 2.6, 1.7 and 1.3 mu m particles were exclusively employed. A fast baseline separation of loratadine and related impurities (R-s,R-min = 2.49) was achieved under the best analytical conditions (i.e. column of 50 mm x 2.1 mm, 1.3 mu m, 10-90% ACN in 5 min, T = 40 degrees C, pH =3, F=0.5 ml/min). This optimal method was successfully tested on columns packed with other particle sizes, namely 1.7 and 2.6 pm, to reduce pressure drop. The selectivities and retentions remained identical, while the peak widths were logically wider, leading to a reduction of peak capacity from 203 to 181 and 159 on the 1.3, 1.7 and 2.6 mu m particles, respectively. On the minimum, the resolution was equal to 1.54 on the 50 mm x 2.1 min, 2.6 pm stationary phase. Next to this, the method was transferred to columns of different lengths, inner diameters and particle sizes (100 mm x 3 mm, 2.6 mu m or 150 mm x 4.6 mm, 5 pm). These columns were used on other LC instruments possessing larger dwell volumes. The modelling software employed for developing the original method was able to calculate the new gradient conditions to be used. The accuracy of prediction was excellent, as the average retention time errors between predicted and observed chromatograms were -0.11% and 0.45% when transferring the method to 100 mm x 3 mm and 150 mm x 4.6 mm columns, respectively. This work proves the usefulness and validity of HPLC modelling software for transferring methods between different instruments, column dimensions and/or flow rates. (c) 2014 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据