4.6 Article

Validated determination of primulasaponins in primula root by a high-performance-thin-layer-chromatography densitometric approach

期刊

出版社

ELSEVIER
DOI: 10.1016/j.jpba.2012.06.040

关键词

HPTLC-densitometry; Primula radix; Primulasaponin I; Primulasaponin II; Saponins; TLC

向作者/读者索取更多资源

A novel HPTLC-densitometric method was developed for separation and quantitation of primulasaponin I and II in different matrices. HPTLC silica gel 60 F254(S), 20 cm x 10 cm, plates with ethyl acetate:water:formic acid (5:1:1 v/v) as the mobile phase were used. Densitometric determinations were performed in reflectance mode at 540 nm after derivatization with vanillin reagent. The method was validated giving rise to a dependable and high throughput procedure well suited to routine applications. Primulasaponins were quantified in the range of 150-450 ng with RSD of repeatability and intermediate precision between 0.8 and 1.4% and accuracy within the acceptance limits. The method was tested on commercial herbal medicinal preparations claiming to contain primula root extract. (C) 2012 Elsevier B.V. All rights reserved.