Evaluation of saliva collection devices for the analysis of steroids, peptides and therapeutic drugs

期刊

出版社

ELSEVIER
DOI: 10.1016/j.jpba.2008.01.033

关键词

saliva collection; steroids; peptides; therapeutic drugs; biomarker; non-invasive

向作者/读者索取更多资源

The recovery of steroids, peptides and therapeutic drugs from commercial saliva collection devices was investigated. Saliva, spiked with defined concentrations of the analytes was applied to the Quantisal (R), three different Salivettes (R), and the Saliva-Collection-System (R) to investigate effects of volume, exposure time and temperature on the recovery. Additionally, saliva was collected from healthy subjects with the same devices. It was found that glucocorticoids can be measured very well from samples obtained with the synthetic fiber Salivettes (R) and the Quantisal (R) (80-100%). For androgens, the Quantisal (R) and the Saliva-Collection-System (R) reached recoveries >80%. The Quantisal (R) and polyester Salivette (R) achieved best recoveries (>80%) for peptides. The results for the cotton Salivette (R) were extremely poor for melatonin, insulin or IL-8 (<20%). The results from the spike-recovery experiments were confirmed by samples collected from healthy volunteers. For most therapeutic drugs the synthetic fiber Salivettes (R) achieved best recoveries of 100 +/- 10%. Longer exposure of saliva on the collection devices must be avoided for most of the analytes, due to their limited stability and increased adsorption. In conclusion, no device is suitable for all of the salivary compounds. Strict pre-analytical precautions must be considered (e.g. immediate processing of the sample) to guarantee reliable analytical results. (C) 2008 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据