4.6 Article

HPLC-UV determination of metformin in human plasma for application in pharmacokinetics and bioequivalence studies

期刊

出版社

ELSEVIER
DOI: 10.1016/j.jpba.2007.10.007

关键词

metformin; human plasma; quantification; reversed-phase chromatography; pharmacokinetics; bioequivalence

向作者/读者索取更多资源

In this study, a simple, rapid and sensitive HPLC method with UV detection is described for determination of metformin in plasma samples from bioequivalence assays. Sample preparation was accomplished through protein precipitation with acetonitrile and chromatographic separation was performed on a reversed-phase phenyl column at 40 degrees C. Mobile phase consisted of a mixture of phosphate buffer and acetonitrile at flow rate of 1.0 ml/min. Wavelength was set at 236 nm. The method was applied to a bioequivalence study of two drug products containing metformin, and allowed determination of metformin at low concentrations with a higher throughput than previously described methods. (c) 2007 Elsevier B.V. All rights reserved.