4.6 Article

Development, validation and transfer into a factory environment of a liquid chromatography tandem mass spectrometry assay for the highly neurotoxic impurity FMTP (4-(4-fluorophenyl)-1-methyl-1,2,3,6-tetrahydropyridine) in paroxetine active pharmaceutical ingredient (API)

期刊

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jpba.2008.08.026

关键词

LC-MS/MS assay; Neurotoxin; Paroxetine; Technology transfer to factory; Sub-ppm impurity

向作者/读者索取更多资源

This work describes the development of a liquid chromatography tandem mass spectrometry (LC-MS/MS) assay for a highly toxic impurity, FMTP (4-(4-fluorophenyl)-1-methyl-1,2,3,6-tetrahydropyridine), in paroxetine active pharmaceutical ingredient (API), followed by the subsequent validation of the methodology and transfer into a global production/quality control environment. The method was developed to achieve a detection limit of 10 ppb mass fraction of FMTP in paroxetine API. An LC-MS/MS method was chosen because it provided the required sensitivity and selectivity with minimal sample preparation. This paper discusses the issues with transferring such complex methodology to a production environment. Linearity, repeatability and reproducibility of the method were demonstrated. This work shows that it is possible using the same approach that would be used for the transfer of any analytical method from R&D to a manufacturing environment. (C) 2008 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据