期刊
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS
卷 46, 期 4, 页码 780-783出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.jpba.2007.11.013
关键词
development; validation; mesylate esters; imatimb mesylate
A gas chromatography-mass spectrometry (GC-MS) method has been developed for the identification and determination of two carcinogenic and genotoxic mesylate esters viz. methyl methanesulfonate (MMS) and ethyl methanesulfonate (EMS) in imatinib mesylate (INM). The method was optimized based on the peak shapes and resolution of MMS and EMS. The method was validated as per International Conference of Harmonization (ICH) guidelines in terms of limits of detection (LOD), limit of quantitation (LOQ), linearity, precision, accuracy, specificity and robustness. The LOD and LOQ values were found to be 0.3 and 1.0 mu g/ml, respectively. The method is linear within the range of 1-15 mu g/ml for both the compounds. These mesylate esters were not found in three different batches of pure and pharmaceutical formulations of INM. (C) 2007 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。