Noise Considerations for PET Quantification Using Maximum and Peak Standardized Uptake Value

期刊

JOURNAL OF NUCLEAR MEDICINE
卷 53, 期 7, 页码 1041-1047

出版社

SOC NUCLEAR MEDICINE INC
DOI: 10.2967/jnumed.111.101733

关键词

PET; standardized uptake value; SUVmax; SUVpeak; treatment monitoring

资金

  1. National Cancer Institute (NCI) [3P30CA 006973-43S2, 1U01CA 140204-01A2]

向作者/读者索取更多资源

In tumor response monitoring studies with F-18-FDG PET, maximum standardized uptake value (SUVmax) is commonly applied as a quantitative metric. Although it has several advantages due to its simplicity of determination, concerns about the influence of image noise on single-pixel SUVmax persist. In this study, we measured aspects of bias and reproducibility associated with SUVmax and the closely related peak SUV (SUVpeak) using real patient data to provide a realistic noise context. Methods: List-mode 3-dimensional PET data were acquired for 15 min over a single bed position in twenty F-18-FDG oncology patients. For each patient, data were sorted so as to form 2 sets of images: respiration-gated images such that each image had statistical quality comparable to a 3 min/bed position scan, and 5 statistically independent (ungated) images of different durations (1, 2, 3, 4, and 5 min). Tumor SUVmax and SUVpeak (12-mm-diameter spheric region of interest positioned so as to maximize the enclosed average) were analyzed in terms of reproducibility and bias. The component of reproducibility due to statistical noise (independent of physiologic and other variables) was measured using paired SUVs from 2 comparable respiration-gated images. Bias was measured as a function of scan duration. Results: Replicate tumor SUV measurements had a within-patient SD of 5.6% +/- 0.9% for SUVmax and 2.5% +/- 0.4% for SUVpeak. SUVmax had average positive biases of 30%, 18%, 12%, 4%, and 5% for the 1-, 2-, 3-, 4-, and 5-min images, respectively. SUVpeak was also biased but to a lesser extent: 11%, 8%, 5%, 1%, and 4% for the 1-, 2-, 3-, 4-, and 5-min images, respectively. Conclusion: The advantages of SUVmax are best exploited when PET images have a high statistical quality. For images with noise properties typically associated with clinical whole-body studies, SUVpeak provides a slightly more robust alternative for assessing the most metabolically active region of tumor.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据