4.2 Article

Effect of moisture uptake on the repeatability of near infrared spectroscopy analyses of ewe milk using the dry extract system for infrared (DESIR) method

期刊

JOURNAL OF NEAR INFRARED SPECTROSCOPY
卷 16, 期 4, 页码 381-388

出版社

SAGE PUBLICATIONS LTD
DOI: 10.1255/jnirs.802

关键词

near infrared spectroscopy; NIR; DESIR; desiccator; repeatability; ewe milk; chemical composition; water absorption

向作者/读者索取更多资源

This paper reports a study carried out to determine the effect of post-filter-drying moisture uptake on the repeatability of NIR spectroscopy analyses of ewe milk using the dry-extract system for infrared (DESIR). The study assessed the spectral repeatability of the method and the effect of moisture uptake on that repeatability, using graphical representations and root mean squared (RMS) error between spectra, as well as multivariate Mahalanobis distances (GH values) and predicted constituent values (protein, fat and dry matter). The spectral repeatability of this method of dried milk analysis using g[ass-fibre fitters was very high, both spectrally per se [RMS MEAN = 3.914 mu log (1/R) and STD = 4.126 mu log (1/1R)] and in terms of NIR spectroscopy prediction of chemical components and of the Mahalanobis distance. Using a single filter per milk sample is thus considered sufficient to guarantee the required accuracy and precision of the DESIR method. Differences between spectra due to the presence or absence of air exposure [STD = 6446 mu log (1/R)] were greater than the method repeatability error. Predictions of protein, fat and dry matter were significantly higher for air-exposed filters, due to moisture uptake; the increase in GH values for air-exposed fitters was highly significant, suggesting that milk samples to be analysed using the DESIR method should be stored in a desiccator after drying and before NIR analysis and not left exposed to air.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据