期刊
JOURNAL OF MICROSCOPY-OXFORD
卷 233, 期 1, 页码 76-83出版社
WILEY-BLACKWELL PUBLISHING, INC
DOI: 10.1111/j.1365-2818.2008.03096.x
关键词
P47; scintillation detector; single electron; response; YAP
类别
The time response of a scintillator-light guide-photomultiplier combination was measured with a time-constant of 3 ns. Single detected electrons were recognizable at the output of the photomultiplier. The distribution of the number of photoelectrons produced by one detected electron and the pulse-height distribution of the photomultiplier output pulses were analysed. Statistical noise computed from these distributions was compared with the noise produced by the dark current of the photomultiplier.