期刊
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS
卷 21, 期 4, 页码 917-925出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/JMEMS.2012.2192910
关键词
Continuous glucose monitoring; interstitial fluid (ISF); microfluidic system; transdermal extraction
类别
资金
- National Natural Science Foundation of China [30800239, 60938002, 61176107]
- Key Projects in the Science and Technology Pillar Program of Tianjin [11ZCKFSY01500]
- Innovation Foundation of Tianjin University
A novel microfluidic system which is fabricated with five polydimethylsiloxane layers for interstitial fluid (ISF) extraction, collection, and measurement toward the application of continuous and real-time glucose monitoring is presented in this paper. The system consists of a micro vacuum generator for ISF transdermal extraction and fluid manipulation, micro chambers for the collection of ISF, micro pneumatic valves for fluid management, and a micro flow sensor for ISF volume measurement. Sequentially controlled by the pneumatic valves, the ISF extraction, collection, and volumetric measurement functions of the system were demonstrated using the stable vacuum generated by the integrated vacuum generator. Through low-frequency ultrasound pretreated full-thickness pig skin, the normal saline solution with different glucose concentrations was transdermally extracted, collected, and measured. The absolute error in the volume measurement of the transdermally extracted ISF analog was less than 0.05 mu L. The microfluidic system makes it possible to realize the clinical application of continuous glucose monitoring based on ISF transdermal extraction technology.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐