JOURNAL OF MICROBIOLOGICAL METHODS
卷 102, 期 -, 页码 55-64出版社
ELSEVIER
DOI: 10.1016/j.mimet.2014.04.013
关键词
Biofilm; Ruggedness; Reproducibility
The MBEC (TM) Physiology & Genetics Assay recently became the first approved ASTM standardized biofilm disinfectant efficacy test method. This report summarizes the results of the standardization process using Pseudomonas aeruginosa biofilms. Initial ruggedness testing of the MBEC method suggests that the assay is rugged (i.e., insensitive) to small changes to the protocol with respect to 4 factors: incubation time of the bacteria (when varied from 16 to 18 h), treatment temperature (20-24 degrees C), sonication duration (25-35 mm), and sonication power (130-480 W). In order to assess the repeatability of MBEC results across multiple tests in the same laboratory and the reproducibility across multiple labs, an 8-lab study was conducted in which 8 concentrations of each of 3 disinfectants (a non-chlorine oxidizer, a phenolic, and a quaternary ammonium compound) were applied to biofilms using the MBEC method. The repeatability and reproducibility of the untreated control biofilms were acceptable, as indicated by small repeatability and reproducibility standard deviations (SD) (0.33 and 0.67 log(10)(CFU/mm(2)), respectively). The repeatability SDs of the biofilm log reductions after application of the 24 concentration and disinfectant combinations ranged from 0.22 to 1.61, and the reproducibility SDs ranged from 0.27 to 1.70. In addition, for each of the 3 disinfectant types considered, the assay was statistically significandy responsive to the increasing treatment concentrations. 2014 Elsevier B.V. All rights reserved.