Relationship Between Measurement Site and Motion Artifacts in Wearable Reflected Photoplethysmography

期刊

JOURNAL OF MEDICAL SYSTEMS
卷 35, 期 5, 页码 969-976

出版社

SPRINGER
DOI: 10.1007/s10916-010-9505-0

关键词

PPG; Motion artifact; Anatomical site; Green light; Infrared light

向作者/读者索取更多资源

Pulse rates obtained from wearable photoplethysmography (PPG) sensors are important for monitoring cardiovascular condition, especially during exercise. However, it is difficult to precisely count pulse rates during exercise because PPG is sensitive to body movement. The artifacts from body movement are caused by a change in the blood volume at the measurement site, in addition to pulsatile changes. Here, we investigated the influence of motion artifact with respect to light source and anatomical sites. In this study, we compared the signal from green-light PPG to that from infrared PPG at different anatomical sites. In these experiments, 11 subjects were asked to either assume a resting position or generate spontaneous motion artifact by jumping and swinging their arm. As a result, pulse rates obtained from green-light PPG showed a higher correlation with the ECG R-R interval as compared to those obtained with infrared. Additionally, the signal from the upper arm showed less artifact than did the peripheral one. Therefore, the green-light PPG may be useful for pulse rate monitoring.