期刊
JOURNAL OF MATERNAL-FETAL & NEONATAL MEDICINE
卷 23, 期 1, 页码 17-22出版社
TAYLOR & FRANCIS LTD
DOI: 10.3109/14767050903156668
关键词
Uterine activity monitoring; electrohysterography
Objectives. Validation of electrohysterography (EHG) as a method for uterine activity monitoring during labour by comparing with intra-uterine pressure catheter (IUPC) recordings. Design. Prospective observational study. Setting. Labour ward in a tertiary centre in the Netherlands. Population. Thirty-two women in labour. Methods. Simultaneous recording of uterine activity with EHG and IUPC for at least 30 min. Main outcome measures. Number of uterine contractions detected by both EHG and IUPC (sensitivity). Number of contractions detected by EHG only [positive predictive value (PPV)]. Correlation between contraction amplitude and duration measured by EHG and IUPC. Results. EHG detects uterine contractions accurately: sensitivity = 94,5% (95%CI: 87.5-100), PPV = 88.3% (95%CI: 76.2-100). The correlation of contractions' duration and amplitude between both methods is r = 0.31 (95%CI: 0.23-0.39) and r = 0.45 (95%CI: 0.38-0.52), respectively. Conclusions. EHG detects uterine contractions accurately during labour but the contraction's characteristics it measures are not directly comparable with that of IUPC.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。