4.3 Article

Nitrogen-doped TiO2 films as drug-binding matrices for the preparation of drug-eluting stents

期刊

JOURNAL OF MATERIALS CHEMISTRY
卷 21, 期 22, 页码 8169-8177

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c0jm03994b

关键词

-

资金

  1. Ministry for Health, Welfare & Family Affairs, Republic of Korea [A084869]

向作者/读者索取更多资源

Nitrogen doped-TiO2 thin film was deposited by plasma enhanced chemical vapour deposition (PECVD) process and its applicability as a drug-binding matrix was investigated in the preparation of drug eluting stents. N-TiO2 film was deposited at 5 W and uniformly covered metallic stent surfaces. Films showed excellent mechanical stability and good adhesion. Alpha lipoic acid, heparin, and abciximab could then be grafted onto the N-TiO2 film through ester bond formation after water plasma modification of the films' surfaces. Drug-grafted N-TiO2 films showed smooth surfaces and good blood compatibility in in vitro platelet adhesion tests. The films maintained uniform coverings, without peeling, even after expansion by a balloon catheter. Grafted-heparin and abciximab could be released continuously for three weeks from drug-grafted N-TiO2 stents in in vitro drug release test. Such immunosuppressive and antiproliferative drug-grafted N-TiO2 film is expected to minimize potential problems related to organic polymers and the risk of direct contact of a bare metal surface with a blood vessel. Hence, the N-TiO2 films prepared in this study are promising alternatives to organic polymers for the preparation of drug-eluting stents.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据