High mass resolution breath analysis using secondary electrospray ionization mass spectrometry assisted by an ion funnel

期刊

JOURNAL OF MASS SPECTROMETRY
卷 47, 期 12, 页码 1571-1575

出版社

WILEY-BLACKWELL
DOI: 10.1002/jms.3118

关键词

secondary electrospray ionization; ion funnel; breath analysis; online mass; spectrometry; FTICR

资金

  1. Sino Swiss Science and Technology Cooperation Program [IZLCZ2 123987]

向作者/读者索取更多资源

In this study, we used secondary electrospray ionization mass spectrometry assisted by an ion funnel (IF) operating at ambient pressure to find compounds in the mass range of 100500 m/z in online breath fingerprinting experiments. In low-resolution experiments conducted on an ion trap instrument, we found that pyridine is present in breath of individuals long after drinking coffee. In high-resolution experiments conducted on a Fourier transform ion cyclotron resonance, we found more than 30 compounds in the mass range of 100500 m/z in analogous online breath experiments. More than a third of these compounds have molecular weights above 200 Daltons and have not been mentioned in previous studies. In low-resolution experiments as well as experiments without the IF, these compounds could not be detected. Copyright (c) 2012 John Wiley & Sons, Ltd.