4.7 Article

A Generalized Strategy for Designing 19F/1H Dual-Frequency MRI Coil for Small Animal Imaging at 4.7 Tesla

期刊

JOURNAL OF MAGNETIC RESONANCE IMAGING
卷 34, 期 1, 页码 245-252

出版社

WILEY
DOI: 10.1002/jmri.22516

关键词

multi-nuclei MRI; dual-frequency RF coil; fluorine imaging; coupled resonator; molecular imaging

资金

  1. NIH [U54 CA119342, R01 HL073646]

向作者/读者索取更多资源

Purpose: To propose and test a universal strategy for building F-19/H-1 dual-frequency RF coil that permits multiple coil geometries. Materials and Methods: The feasibility to design F-19/H-1 dual-frequency RF coil based on coupled resonator model was investigated. A series capacitive matching network enables robust impedance matching for both harmonic oscillating modes of the coupled resonator. Two typical designs of F-19/H-1 volume coils (birdcage and saddle) at 4.7T were implemented and evaluated with electrical bench test and in vivo F-19/H-1 dual-nuclei imaging. Results: For various combinations of internal resistances of the sample coil and secondary resonator, numerical solutions for the tunable capacitors to optimize impedance matching were obtained using a root-seeking program. Identical and homogeneous B1 field distribution at F-19 and H-1 frequencies were observed in bench test and phantom image. Finally, in vivo mouse imaging confirmed the sensitivity and homogeneity of the F-19/H-1 dual-frequency coil design. Conclusion: A generalized strategy for designing F-19/H-1 dual-frequency coils based on the coupled resonator approach was developed and validated. A unique feature of this design is that it preserves the B1 field homogeneity of the RF coil at both resonant frequencies. Thus it minimizes the susceptibility effect on image co-registration.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据