期刊
JOURNAL OF MAGNETIC RESONANCE IMAGING
卷 28, 期 1, 页码 13-20出版社
JOHN WILEY & SONS INC
DOI: 10.1002/jmri.21421
关键词
time-of-flight MR angiography; 3T MRI; aneurysm; platinum coil; short echo time; high-spatial-resolution MR angiography; phantom
Purpose: To analyze the influence of matrix and echo time (TE) of three-dimensional time-of-flight (3D TOF) magnetic resonance angiography (MRA) on the depiction of residual flow in aneurysms wtih platinum coils at 1.5T and 3T Materials and Methods: A simulated intracranial aneurysm of the vascular phantom was loosely packed to maintain the patency of some residual aneurysmal lumen with platinum coils and connected to an electromagnetic flow pump with pulsatile flow. MRAs were obtained altering the matrix and TE of 3D TOF sequences at 1.5T and 3T. Results: The increased spatial resolution and the shorter TE offered better image quality at 3T. For the depiction of an aneurysm remnant, the high-spatial-resolution 3T MRA (matrix size of 384 x 224 and 512 x 256) with a short TE of <= 3.3 msec were superior to the 1.5T MRA obtained with any sequences. Conclusion: 3T MRA is superior to 1.5T MRA for the assessment of aneurysm embolized with platinum coils: the combination of the 512 x 256 matrix and short TE (3.3 msec or less) seems feasible at 3T.